Technical parameters
| Max Voltage | 150KV | 
| Maximum X-ray field coverage range | 430mm×430mm (SID=100cm) | 
| The average brightness of the light field | >160 lux | 
| Edge contrast ratio | >4:1 | 
| Power supply requirement of projection lamp | 24V AC/150W | 
| Bright X-ray field duration for Once | 30S | 
| Distance from Focal Spot of X-ray tube to the mount plane of collimator SID (mm)(optional) | 60 | 
| Filtration(Inherent) 75kV | 1mmAl | 
| Filtration(Additional) | The outer optional | 
| Control method | Electric/ Manual | 
| Drive motor | Stepper Motor | 
| Motor Control | TTL Level signals | 
| Position detection | Potentiometer | 
| Input power | AC24V/DC24V | 
| (SID)Measuring tape | optional | 
| Center laser instructions | optional | 
| Dimension(mm)(W×L×H) | 185×230×145 | 
| Weight(Kg) | 7.2 | 
Applications
This x-ray collimator is suitable for tube voltage 150kV, DR digital and common X-ray diagnostic equipmentMain technical parameters and indicators
X-ray leakage: <1mGy/h (150kV, 4mA) The distance from the focus of the X-ray tube to the beam limiter mounting surface: 60mm (can be adjusted according to different tubes) Maximum irradiation field: 43cmX43cm (SID=1m) Minimum irradiation field: <5cmX5cm (SID=1m)  Visible light field brightness: >140lux (SID=1m) Light field consistency: <2%@SID Inherent filtration: 1mmAl/75kV Light field power input: 24VAC/50W or DC24V/1A  Electrical control power input: 24VDC/2A CAN bus protocol: CAN2.0A  Dimensions: 186mm×230mm×145mm (length×width×height) Weight: 7.2kg Optional: External additional filter Special electrical interface Special tube ball interface